Thank you for taking interest in Daniel Consultants Inc part number N0-059, which is listed with the NSN 6630-01-364-8555 and a product description of “Analyzer Blood Gas.” In order to receive procurement options on this particular listing, we ask that you fill out and submit the form provided below with as much information as you can concerning your needs. By supplying our staff with key details on your desired quantity, pricing target, and expedited fulfillment timeframe, we can respond with a quote that has been specifically tailored to your unique specifications. We also ask that you make us aware of any restrictions that you may have as soon as possible, such as if you have a time constraint or are seeking unlisted items.
Note : We Will Not Share Your Information To Any Third Parties.
Alongside part number N0-059, you are encouraged to continue exploring our collection curated catalogs to find additional items that can accompany this Analyzer Blood Gas part with the NSN 6630-01-364-8555. For example, you can browse other Daniel Consultants Inc components that were produced under CAGE Code 0NTK1 or listings that belong to FSC 6630 Chemical Analysis Instruments with ease on our website, with search tools provided for quick location. Furthermore, our staff is just a call or email away from assisting you with any inquiries or requests, so be sure to get started on the quotation process for part number N0-059 and other desired offerings on Aerospace Spare.
MRC | Criteria | Characteristic |
---|---|---|
DPTH | Depth | 18.300 INCHES |
ELEC | Voltage In Volts | 120.0 AND 230.0 |
MRC | Decoded Requirement | Clear Text Reply |
CBBL | Features Provided | DIGITAL DISPLAY SCREEN AND INTEGRAL PRINTER AND SELF-CALIBRATION |
CXCY | Part Name Assigned By Controlling Agency | BLOOD GAS ANALYZER 120/230V 50/60 HZ AC: |
ACDC | Current Type | AC |
CYFV | Method Of Analysis | POLAROGRAPHIC AND POTENTIOMETRIC/ION SELECTIVE ELECTRODE |
ACZB | Frequency Rating | 50.0 HERTZ NOMINAL OR 60.0 HERTZ NOMINAL |
CYFY | Analysis Test/Measurement Range | PO2 0 TO 600 MMHG; PH 6.8 TO 7.8; PCO2 8 TO 99 MMHG; K 0.1 TO 9.9 MMOL/L; NA 100 TO 200 MMOL/L; CA 0.25 TO 4.25 MMOL/L; HCT 10 TO 75 PCT; HC03 3.0 60 MMOL/L TCO2 3.0 TO 60 MMOL/L; BASE EXCESS -30 TO +30 MMOL/L; O2 SAT 30 TO 90 PCT. |
CYHM | Test Sample/Specimen Type | WHOLE BLOOD |
ABRN | Height | 8.400 INCHES |
WDTH | Width | 9.100 INCHES |
AQXY | Test Type For Which Designed | BLOOD GASES AND ELECTROLYTES,HC03,HCT,BASE EXCESS AND 02 SAT |
CYFZ | Test Sample/Specimen Size | 0.5 ML |
FEAT | Special Features | UTILIZES A DISPOSABLE CARTRIDGE/TEST PACK THAT CONTAINS ALL ELECTRODES, CALIBRATING SOLUTIONS AND GASES PLUS A WASTE BAG FOR PROCESSED BLOOD, CALIBRATING AND QC SOLUTIONS. THE CARTRIDGE/TEST PACK SHALL ELIMINATE THE NEED FOR GAS TANKS EXTERNAL SOLUTIONS, AND ELECTRODE MAINTENANCE. SHALL BE PRE-PROGRAMED TO AUTOMATICALLY PERFORM TWO-POINT CALIBRATIONS EVERY HOUR AND ONE-POINT CALIBRATIONS EVERY 90 SECONDS. CAPABLE OF AUTOMATIC ASPIRATION OF THE SAMPLE AND TO PROCESS, DISPLAY, AND PRINT THE RESULTS IN LESS THAN 2 MINUTES. SHALL BE SUITABLE FOR THE FIELD ENVIRONMENT. |